Specification
A. Principles of Measurement
Blood Cells are counted and sized by Electrical Impedance Method. Hemoglobin is determined by Colorimetry Method.
B. Sampling Features
1.Sample Volume
Venous Blood Mode: 10μL Venous Blood
Prediluted Mode: 20μL Capillary Blood
2.Reagent Volumes Required for Single Sample
Diluent: 30mL
Rinse: 8mL
Lyse: 0.35 mL
3.Venous Mode
WBC/HGB: 1:300
RBC/PLT: 1:44600
A. Principles of Measurement
Blood Cells are counted and sized by Electrical Impedance Method. Hemoglobin is determined by Colorimetry Method.
B. Sampling Features
1.Sample Volume
Venous Blood Mode: 10μL Venous Blood
Prediluted Mode: 20μL Capillary Blood
2.Reagent Volumes Required for Single Sample
Diluent: 30mL
Rinse: 8mL
Lyse: 0.35 mL
3.Venous Mode
WBC/HGB: 1:300
RBC/PLT: 1:44600
4. Prediluted
Mode
WBC/HGB: 1:355
RBC/PLT: 1:44500
5.Cell Counting Aperture Size
WBC/RBC/PLT: 80μm
5. Display
Liquid Crystal Display (LCD),resolution:640×480
C. Indicator
20 Parameters and 3 Histograms Display
Alarm Indicator
Working Status(including Power) Indicator Light Alarm
D. Interface
One power input socket
One printer port
Two RS-232 ports
Two USB interfaces
Two PS/2 interfaces
E. Recorder
Rapid Thermal Recorder
Recording width: 48mm
Paper width: 57.5mm
RBC/PLT: 1:44500
5.Cell Counting Aperture Size
WBC/RBC/PLT: 80μm
5. Display
Liquid Crystal Display (LCD),resolution:640×480
C. Indicator
20 Parameters and 3 Histograms Display
Alarm Indicator
Working Status(including Power) Indicator Light Alarm
D. Interface
One power input socket
One printer port
Two RS-232 ports
Two USB interfaces
Two PS/2 interfaces
E. Recorder
Rapid Thermal Recorder
Recording width: 48mm
Paper width: 57.5mm