This is a dedicated system device for respiration salvaging and revival, The main frame choose the principle of high-frequency ventilation, it is widely used for clinical salvaging and revival also is a dedicated device used for respiration salvaging on first-aid car.
Features
1. An opening system is designed to
avoid leakage of airway, When ventilating, the intubation may without casing
bag.
Little tube may be used for ventilation when resistance of expiration is
small, so that mucosa won't be hurt easily.
2. With opening airway, suction and ventilation
can be done at the same time, so good effectiveness of oxygen
combination can
be shown without inspiration and expiration synchronizing.
3. As ventilating, lower airway pressure is good
to increase blood discharge and decrease encephal pressure.
4. Suction device, it driving negative pressure
is adjustable.
5. Accumulator jar, it can be used continuously
for not less than four hours after being charging.
6. The system disposes completely with
laryngoscope, inlet intubation, catheter of tracheotomy,
nasal prongs tube, air
tube and so on.
Main
Specifications
1. Output power: ≤65VA Power
supply: AC220V+22V 50HZ+1HZ
2. Frequency:10~150 breaths/min
3. I/E ratio:1:1, 1:1.5~1:3 and inverse ratio
4. SIGH:6
breaths/min * CPAP: Not less than
1.5KPa
5. Suction negative pressure:-0.01~0MPa
6. Driving Pressure:0~0.15MPa Relative humidity :
≤80% Temp.: 5-40%
7. Pressure Regulator Limit:0.30MPa
8. Constant volume ventilation:
T=10min,I/E=1:1.5
Adults: f(100):f(20)=7:3 tidal volume=550ml/time Driving
pressure=0.06MPa
Children: f(100):f(20)=7:3 tidal volume=350ml/time Driving
pressure=0.04MPa
9. First-aid: 1) adult T=10min,
F(100)=7min, F(20)=3min, Tidal=550ml/time, Cycle
2) Child T=10min, F(150)=7min, F(30)=3min, Tidal=350ml/time, Cycle
10. Storage battery:4Ah,Automatic charge
when 12V
11. Most jetting volume: ≥1400ML/time
12. Oxygen concentration: 35-90% adjustable
13. Humidifying ratio: 0-15ml/hour
14. Airway pressure display: 0-3Kpa
15. Alarm: Alarm when battery pressure is lower
than 11V;Alarm when the oxygen pressure is
lower than 0.01MPa;Alarm with sound
and light when no gas
16. Ventilation Mode: HFJV, CPAP, HFJV+SIGH(Sigh),
Constant volume ventilation.