Hemodialysis Machine TNC-4000



Forms of Therapy: 
Hemodialysis, Online Hemofiltration (Online HF), Online Hemodiafiltration (Online HDF),Isolated Ultrafiltration, Sequential Ultrafiltration, Hemoperfusion, Plasma exchange (optional).

Technical Features:
1. The Multi-Functional Machine Offers Customized Curing Mode For Patients. 
2. The Advanced Volume Balance Control System Enables   Accurate 
    Control of Ultrafiltration. 
3. Double-Stage Intoxication Filtration Ensuring That The Substitution 
    Fluid Meets AAMI Standards. 
4. Reliable System and Safe Treatment Ensured By Pressure Detectors, 
    Conductivity Detectors, Air Detectors and Blood Leakage Detectors.




Function Features: 

1. Adjustable sodium And UF profiling Enable Customized Treatment For Patients. 
2. Convenient Process for Sequential Ultrafiltration and Dialysis. 
3. High Volume Hemofiltration (HVHF) Is Available 
4. Kt/V Calculation Function Can Be Used For Evaluating the Sufficiency of Dialysis via the Urea Kinetic 
    Model Volume Balance System. 
5. Isolated Hemoperfusion Mode, Decrease the Normal Hemodiafiltration Device Self-Detecting 
    Time And The Dialysate Waste. 
6. The Adaptive Regulation and Control System Effectively Avoids Various Complications. 
    And Can Clear Middle and Small Molecules Toxin 
7. Automatic Disinfestation and Decalcifilcation by Many Kinds of Disinfectant Liquids and Picking Solutions. 
8. Back-Up Power Supply For 30 Min After Electricity Off.

 Technical Data:
Dimensions (W x H x L)
850 × 840×1660 MM
Gross Weight
Aprox. 140 Kg
Power Supply
AC220V, 50~60HZ
Connected Load
2000VA
Water Supply
Pressure Range :0~8.0BAR, Temperature Range:0~35
Venous Pressure Working Range
-700~+800mmhg, Accuracy±10mmhg
Arterial Pressure Working Range
-700~+800mmhg, Accuracy±10mmhg
Transmembrane Pressure Working Range
-700~+800mmhg, Accuracy±10mmhg
Blood Flow Range
0~700ml/Min
Dialysate Flow Range
0~300~500~800ml/Min (Adjustable)
Substitution Fluid Flow
0~25L/H
Dialysate Temperature Working Range
30~41,Accuracy±0.1
Heparin Pump Working Range
0~10 ml/H,Accuracy±0.1ml/H
Conductivity Working Range
10~20 ms/cm Accuracy±0.1ms/cm
Air Detector
Detecting≥0.02ML/Single Bubble Detecting
Blood Leak Detector
Detecting blood leak ≥0.35mL/min (HCT 32%)
Hot Rinse
85
Clean/ Disinfection
Decalcification: Rinse, citric acid
Thermal Disinfection: Hot water or citric acid
Chemical Disinfection: Sodium hypochlorite orperacetic acid