1. The high-precision evaporator has the
stable output concentration with pressure,
temperature, and flow automatic
compensation; Choose one among Enflurane,
Isoflurane and Sevoflurane.
2. Reliable gas Power transmission system;
3. Reasonable economic and multiple
technology configuration, the user s may choose flexibly;
4. Modular product design can be upgrade
according to users’ need
Main Technical Parameters
Part of Anesthesia Machine
1.
Console
high-strength engineering plastic rack, light, beautiful and corrosion
resistance
2.
Scope adult, child
3.
Gas source O2:0.3~0.5MPa
4.
Flow meter O2:0.1~2L/min 2.5~10L/min
5.
Rapid oxygen flow 35~75L/min
6.
Low oxygen pressure
alarm it will be alarmed when the oxygen pressure < 0.2MPa
7.
Vaporizer
It has the function of pressure, temperature, and flow automatic compensation. the regulation range of evaporator concentration is 0~5vol%. Choose one among Enflurane, Isoflurane and Sevoflurane for a matching according to the demands of customer.
It has the function of pressure, temperature, and flow automatic compensation. the regulation range of evaporator concentration is 0~5vol%. Choose one among Enflurane, Isoflurane and Sevoflurane for a matching according to the demands of customer.
8.
Respiratory circuit
work mode all close. semi-close. semi-open pressure release 1~4KPa
9.
Bellows breathing
adults and children bellows tidal volume range:0~1500ml
Part of
Ventilator
1. Display mode high brightness LED display
2. Ventilation mode IPPV
3. Ventilation parameters
Tidal volume 50~1500ml
Frequency 6~60 bpm
Breath /expiratory ratio 2:1~1:3
Airway pressure range -2~10 Kpa
Tidal volume 50~1500ml
Frequency 6~60 bpm
Breath /expiratory ratio 2:1~1:3
Airway pressure range -2~10 Kpa
4. Ventilation monitoring parameters
Tidal volume. ventilation. respiratory rate. airway pressure
Breath /expiratory ratio.
Tidal volume. ventilation. respiratory rate. airway pressure
Breath /expiratory ratio.
5. Security Alert System
Alarm minute ventilation
Fixed ceiling value 18L/min
Fixed threshold value 1~10L/min
Alarm minute ventilation
Fixed ceiling value 18L/min
Fixed threshold value 1~10L/min
6. Asphyxia alarm
The machine will give sound and light alarm if there is no tidal volume input in 15s
Power failure alarm
The machine will give sound and light alarm if there is no tidal volume input in 15s
Power failure alarm