TNC-2207 Ventilator is a model of multi-purpose life-support machine. This machine uses the air operated electric control, the microcomputer technology. Selects imports the primary device to guarantee the equipment movement stably reliable. Colored (TFT) the screen demonstrates each kind of examination and the establishment parameter, which is a section of ideal breath treatment plant. It is suitable in all different levels of hospitals for internal medicine of surgery, gynecology, pedology, first aid, ICU and so on which are administrative offices equipments
Main technical parameters
1. Display mode 10.4 inch high-definition TFT
color LCDdisplay
2. Ventilation mode
IPPV SIPPV IMV SIMV PCV PEEP/CPAP SIGH
3. Ventilation parameters
Tidal volume 50~1500ml
Frequency 4~80bpm
I:E 4:1~1:4
SIMV Frequency 2~12 bpm
Inspiratory pressure trigger -0.1~-1.0KPa
PEEP 0.1~1.0KPa
Pressure control 1~6KPa
Inspiratory pause 0~50% inspiratory time
Sigh control breathing in a deep breath every 10~120 times/min. Inspiratory time is 1.5 times
to the settings.
4. Ventilation monitoring
Tidal volume,Ventilation capacity,IPPV frequency,Breathing
frequency,Inspiratory ime/expiratory time ratio,Peak value
of airway
pressure,Pressure-time wave,Flow rate-time wave,
Breathing frequency ,End-expiratory positive pressure ,
Inspiratory trigger pressure
5. Safety alarm system
Airway pressure alarm
Upper limit setting range 0~6.0KPa
Low limit setting range 0~5.0KPa
Tidal volume
Upper limit setting range 0~2000ml
Low limit setting range 0~1800ml
Oxygen concentration alarm
Upper limit setting range 21%~100%
Low limit setting range 0%~80%
Asphyxia alarm
The machine will give sound and light alarm if there is no
Tidal volume in15s
Power failure alarm
6. Power AC 220V±10% 50Hz±1Hz UPS,Battery (choose and match)