1. This machine uses in respiratory control,respiratory intensive care , 5.1 inch
high clear display, and real-time reflection respiratory system condition.
2. The high-precision evaporator has the stable output concentration
with pressure, temperature, and flow automatic compensation;
Choose one among Enflurane, isoflurane and sevoflurane for a matching.
3. Reliable gas Power transmission system;
4. Reasonable economic and multiple technology configuration, the user s
may choose flexibly;
5. Modular product design According to customer’s needs upgrade.
Main Technical Parameters
Part of Anesthesia Machine
1.
Console high-strength
engineering plastic rack,light,beautiful and corrosion resistance
2.
Scope adults,
children
3.
Gas source O2、N2O:0.3~0.5MPa
4.
Flow meter O2: 0.1~10L/min
N2O : o.1~10L/min
N2O : o.1~10L/min
5.
O2 N2O linkage and nitrous oxide Truncation device
When use nitrous oxide, oxygen concentration>25%; when the oxygen pressure<0.2Kpa, the flow of
nitrous oxide would be cut off.
When use nitrous oxide, oxygen concentration>25%; when the oxygen pressure<0.2Kpa, the flow of
nitrous oxide would be cut off.
6.
Rapid oxygen flow 35~75L/min
7.
Vaporizer
It has the function of pressure, temperature, and flow automatic compensation.
The regulation range of evaporator concentration is 0~5vol%. Choose one among Enflurane,
Isoflurane and Sevoflurane for a matching according to the demands of customer.
It has the function of pressure, temperature, and flow automatic compensation.
The regulation range of evaporator concentration is 0~5vol%. Choose one among Enflurane,
Isoflurane and Sevoflurane for a matching according to the demands of customer.
8.
Low oxygen pressure
alarm It will be alarmed when the oxygen pressure < 0.2MPa
9.
Respiratory circuit Work mode all close. semi-close. semi-open pressure release 1~4KPa
10. Bellows breathing
Adults and children bellows tidal volume range:0~1500ml
Adults and children bellows tidal volume range:0~1500ml
Part of
Ventilator
1.
Display mode
5.1-inch high clear display,
2.
Ventilation mode
IPPV SIPPV PLV PEEP MANU (monitor the tidal volume. ventilation. respiratory rate. ratio. peak pressure when use hand.)
IPPV SIPPV PLV PEEP MANU (monitor the tidal volume. ventilation. respiratory rate. ratio. peak pressure when use hand.)
3.
Ventilation
parameters
tidal volume 50~1500ml
frequency 6~60bpm
I:E 2:1~1:4
Trigger sensitivity -1.0~-0.2Kpa
PEEP 0.2~1.0Kpa
pressure control 1~6.0Kpa
tidal volume 50~1500ml
frequency 6~60bpm
I:E 2:1~1:4
Trigger sensitivity -1.0~-0.2Kpa
PEEP 0.2~1.0Kpa
pressure control 1~6.0Kpa
4.
Ventilation
monitoring parameters
tidal volume. ventilation. IPPV ratio. total respiratory rate. breath /expiratory ratio. peak pressure . pressure control. airway pressure wave. breath pressure trigger. PEEP minimum pressure.
tidal volume. ventilation. IPPV ratio. total respiratory rate. breath /expiratory ratio. peak pressure . pressure control. airway pressure wave. breath pressure trigger. PEEP minimum pressure.
5.
Monitoring oxygen
concentration 21%~100%(choose
and match)
6.
Security Alert System
Oxygen Concentration alarm (choose and match)
upper limit setting range 50%~100%
low limit setting range 15%~50%
Oxygen Concentration alarm (choose and match)
upper limit setting range 50%~100%
low limit setting range 15%~50%
7.
Airway pressure
alarm
upper limit setting range 1.5~6.0Kpa
low limit setting range 0.2~3.0Kpa
upper limit setting range 1.5~6.0Kpa
low limit setting range 0.2~3.0Kpa
8.
Alarm minute
ventilation
upper limit setting range 2~30L/min
low limit setting range 1~12L/min
upper limit setting range 2~30L/min
low limit setting range 1~12L/min
9.
Asphyxia alarm The
machine will give sound and light alarm if there is no tidal volume input in
15s
Power failure alarm
Power failure alarm
10. Power AC 220V±10% 50Hz±1Hz UPS. Battery. (choose and
match).