TNC-2103 Anesthesia Machine with Two Vaporizers Two Gas























Main Technical Parameters
Part of Anesthesia Machine

1. Console high-strength engineering plastic rack, light, beautiful and 
    corrosion resistance
2. Scope adult, child
3. Gas source O2N2O:0.30.5MPa
4. Flow meter O2: 0.110L/min 1.110L/min
    N2O: 0.1
10L/min 1.110L/min
5. O2 N2O oxide linkage and nitrous oxide Truncation device
    When use nitrous oxide, oxygen concentration>25%; when 

     the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
6.  Rapid oxygen flow 3575L/min
7.  Vaporizer it has the function of pressure,temperature, and flow automatic compensation.
     The regulation range of evaporator concentration is 05vol%. Choose two among enflurane, 
      isoflurane and sevoflurane for a matching according to the demands of customer.
8.   Low oxygen pressure alarm It will be alarmed when the oxygen pressure < 0.2MPa
9.   Respiratory circuit work mode all close. semi-close. semi-open APL: 0.57KPa
10. Bellows breathing adults and child bellows tidal volume range:01500ml child bellows tidal 
      volume range:0300ml adults children bellow children loop are arbitrary choice for users.                                                                                                            

Part of Ventilator
1.      Display mode 5.7-inch high clear display,
2.      Ventilation mode IPPV SIPPV SIMV PEEP MANUAL SIGH
3.      Ventilation parameters
  Tidal volume 50
1500ml
  Frequency 4
60bpm
  I:E 2:1
1:8
 Trigger sensitivity -1.0
-0.1Kpa
 PEEP 0
1.0Kpa
 Pressure Control 1
6Kpa
 SIGH 1.5 breath times
4.     Ventilation monitoring parameters
Tidal volume. ventilation. IPPV ratio. breath /expiratory
Ratio. peak pressure . pressure-time wave. velocity-time
Wave. total respiratory rate. airway pressure wave. PEEP.breath pressure trigger.
5.      Monitoring oxygen concentration 21%100%(choose and match)
6.      Security Alert System (choose and match)
Oxygen Concentration alarm
Upper limit setting range 50%
100%
Low limit setting range 15%
50%
7.      Airway pressure alarm
Upper limit setting range 1
6Kpa
Low limit setting range 0.4
2Kpa
8.      Alarm minute ventilation
Upper limit setting range 3.0
30L/min
Low limit setting range 1
10L/min
9.      Asphyxia alarm the machine will give sound andlight alarm if there is no tidal volume input in 15s
Power failure alarm
10.  Power AC 220V±10% 50Hz±1Hz UPS. Battery. (choose and match)
11.  Manipulators (choose and match)