TNC-2401 Infant Radiant Warmer



Feature:
1. Microprocessor based servo controlled temperature system.
2. Digital display to indicate temperature
3. Protection when temperature probe in low temperature.
4. Alarms functions for safety includes:
    Power failure temperature deviation
5. Temperature probe failure over temperature 
6. Have independent the second heat cuts off , the system is more safe.

7. Heater is electrically cut off if air /skin temperature inside incubator exceed 40°C.



Technical specification:

1. Electrical requirement: 220V±10%/50±2%
2. Power supply: ≤1000VA
3. Temperature control range: 32℃~38.5℃
4. Temperature probe accuracy: ≤0.3℃
5. Mattress temperature uniformity: ≤2.0℃
6. Warm-up time: ≤45min

Operation condition:

1. Environment temperature: 18℃~30℃
2. Environment relative humidity: 30%~75%RH 
3. Environment air velocity of flow: <0.3m/s

Package:
Gross weight: 100kg
Net weight: 70kg
Measurement: 1100*780*860mm