TNC-2402 Infant Radiant Warmer



 Feature:
1. Microprocessor based servo controlled temperature system.
2. Digital display to indicate temperature
3. Protection when temperature probe in low temperature.
4. Alarms functions for safety includes:
5. Power failure temperature deviation
6. Temperature probe failure over temperature 
7. Have independent the second heat cuts off , the system is more safe.
8. Heater is electrically cut off if air /skin temperature inside incubator exceed 40°C.




Technical specification:
 1. Electrical requirement: 220V±10%/50±2%
 2. Power supply: ≤1000VA
 3. Temperature control range: 32~38.5
 4. Temperature probe accuracy: ≤0.3
 5. Mattress temperature uniformity: ≤2.0
 6. Warm-up time: ≤45min

Operation condition:
 1. Environment temperature: 18~30
 2. Environment relative humidity: 30%~75%RH 
 3. Environment air velocity of flow: <0.3m/s

Package
 1. Gross weight: 100kg
 2. Net weight: 70kg
 3. Measurement: 1100*780*860mm

Related product you might see: