TNC-4103 200mA 100KV (125KV) MEDICAL X-RAY Machine





 Usage
 Roentgenoscopy
 Digestive track spot film photography 
 General photography 
 Filter-ray device photography 









Features
1. Single bed and single bulb 
2. Design according to the electric mechanism of middle-size machine. The function is full and reasonable. 
3. The high-tension system is of march-past structure. Equipped with full-wave rectification high-voltage 
    generator and independent X-ray tube unit, this machine is with large output power and operation is reliable and stable. 
4. Supply voltage V, roentgenoscopy, photography KV and step-less adjustable. 
5. Equipped with the manostat for filament of X-ray tube and the space-charge compensator as to ensure 
    the stable and accurate output. 
6. Equipped with the capacity limitation and protection circuit and KvmAS interlock limitation and protection. 
7. Adopt the digital circuit timer, grade according to R10 priority coefficient and can control time exactly 
8. Examining bed is of electric back rotation and is equipped with vibrating filter-ray device. 
9. The spot film device of luminescent screen can hold the film jacket of various sizes and can solenoid valve locking. 
10. The upright post is of top and bottom track. It moves lightly.


Main specifications 
ITEM
CONTENT
INDEX
Roentgenoscopy
voltage
50-95KV Be adjustable continuously
current
0.5~5mA
Photography
voltage
50-100KV,50-125KV Be adjustable continuously
current
small focus 50mA; big focus 50, 100, 150, 200mA five grades together
spot film
150mA (fixed)
time
0.04~6.3s,23 grades together
High-Voltage generator
capacity
20KVA(instantaneous)
Max. Dc output voltage
100KV 125KV
Max. Dc output current
200mA
X-Ray tube
model
XD51-20.40/100 XD51-20.40/125
focus
1X1mm 2X2mm
Examining Bed
rotary angle
+90°- 0° - -10°
specification of spot film(stomach and intestines photography)
lengthwise 500mm
transverse ± 80mm
pressure down 300mm

Examining bed
specification of spot film
127X177.8mm(5”X7”)
203.2X254mm(8”X10”)
254X304.8mm(10”X12”)
279.4X355.6mm(11”X14”)
filter-ray device
grid characteristic
N28
Convergence distance
fo 100cm
Grid ratio
r8
distance between bed face and film
75mm
overrunning governor
manual single blade
Upright post
adjusting scope (distance between top and bottom track)
2800-3400mm
move lengthwise
(along top and bottom track)
3100mm
move transverse
(along horizontal arm)
800mm(from bulb focus to the centre of upright post)
move up and down
(along upright post)
930-1830mm(from bulb focus to the centre of upright post)
Supply power
voltage
380V/220V
frequency
50Hz ± 0.5Hz
inner resistance
380V:0.9 Ω ,220V:0.3 Ω