Usage
Fluoroscopy
Gastrointestinal radiography
General radiography
Ray-filter-ray radiography
Features
1. Two tables and two tubes X-ray machine is separate and special, with fluoroscopy, gastrointestinal photography,
general photography, filter-ray photography.
2. Three
parameters kV mA and s are sectional interlocked protection in capacity
Both fluoroscopy and photography are equipment
with limited time protection fittings.
3. Equipment
with protection of the manostat for filament of X-ray tube, Space charge
compensating, Rotary
anode starting and overrunning–temperature protection
Supply power V, fluoroscopy kV, photography kV
and motor adjusts continuously.
4. Equipment
with the electric diagnostic bed, fluorescent photography device and
electromagnetism locks
The photography table can be used
professionally, operated expediently, the ground-track is selected
fittings
when using.
Main specifications
ITEM
|
CONTENT
|
INDEX
|
Outer
power
|
Capacity
|
30KVA
|
Voltage,
frequency
|
380V/220V±10%
50Hz±1Hz
|
|
Inner
resistance
|
0.9Ω
at 380V; 0.2Ω at 220V
|
|
Supply
control
|
Scope
|
380V±10%,
Be adjustable continuously or 220V±10%, Be adjustable continuously
|
High-voltage generator |
Max-capacity
|
25KVA(instantaneous)
|
Max
Dc output voltage
|
125KVP
|
|
Max
Dc output current
|
300mA(instantaneous)
|
|
Fluoroscopy
|
Tube
voltage
|
45-100KV
Electric Be adjustable continuously
|
Tube
current
|
0.5-5mA
Be adjustable continuously
|
|
Gastrointestinal
radiography
|
Tube
voltage
|
50-110KVP
|
Tube
current
|
150mA(fixed)
|
|
Fluoroscopy
Gastrointestinal radiography X-Ray tube
|
Focus
|
small1×1mm
; big 2×2mm
|
Model
|
XD51-20、40/125
|
|
General
radiography、filter-ray radiography
|
Tube
voltage
|
50-125KVP
|
Tube
current
|
Small
focus 50,100mA ; big focus 50,100,150,200mA
|
|
Radiography
X-Ray tube
|
Focus
|
Small
focus:1×1mm big focus: 2×2mm
|
Model
|
XD51-20,40/125
|
|
Diagnostic
bed
|
Rotary
scope
|
90°-0°--10°
|
Bed
L×W×H
|
1960×702×820mm
|
|
Fluoroscopy
Gastrointestinal radiography equipment
|
Horizontal
|
About
±90mm
|
Up
and down
|
580mm
|
|
Lock
mode
|
electromagnetic
lock
|
|
Light
control
|
Manual
single-vane limiter
|
|
Photograph
specification
|
128×179mm(5”×7”)
in two
|
|
205×254mm(8”×10”)
in two
|
||
254×305mm(10”×12”)
in two
|
||
279×355mm(11”×14”)
in four
|
||
Radiography
bed
|
L×W×H
(mm)
|
2000×710×700mm
|
Lengthwise
|
700mm
|
|
Vibration
Ray-filter device
|
Lengthwise
distance >=550mm
|
|
Grid
density N28
|
||
grid
ratio r8
|
||
Convergence
distance f0100cm
|
||
X-Ray
tube unit-support
|
Lengthwise
|
2100mm
|
Up
and down
|
325-1825mm
|
|
Tube
turning around the shaft of arm
|
180°
|
|
Turning
around the shaft of tube
|
-10°-90°-120°
|