TNC-4106 300mA Medical X-ray Camera






Usage : General Photograph Ray-filter Photograph











Features 
1. Single bed and single bulb 
2. Design according to the electric mechanism of middle-size machine. The function is full and reasonable. 
3. The high-tension system is of march-past structure. Equipped with full-wave rectification high-voltage generator and
    independent X-ray tube unit, this machine is with large output power and operation is reliable and stable. 
4. Supply voltage V, roentgenoscopy, photography KV and step-less adjustable. 
5. Equipped with the manostat for filament of X-ray tube and the space-charge compensator as to ensure
    the stable and accurate output. 
6. Equipped with the capacity limitation and protection circuit and KvmAS interlock limitation and protection. 
7. Adopt the digital circuit timer, grade according to R10 priority coefficient and can control time exactly 
8. High-voltage primary uses the zero controlled circuit of silicon-controlled rectifier of large power. 
9. Photographic bed can move in length and breadth 
10. The photographic bed, pillar and vibrating ray-filter are in a whole without top and bottom track.

Main specifications
ITEM
CONTENT
INDEX
Supply power
capacity
Two-phase 30KVA
voltage frequency
380V/220V ± 10% 50Hz ± 0.5Hz
inner resistance
0.75 Ω at 380V, 0.25 Ω at 220V
Power supply
Adjustable range
380V,380V ± 10% ,220V ,220V ± 10%being adjustable continuously
Photography
voltage
50-125KV Be adjustable continuously
current
small focus 50mA 100mA, big focus 50,100,200, 300mA
time
0.04-6.3s,23 grades together according to R10 coefficient
High-Voltage generator
capacity
20KVA(instantaneous)
Max. Dc output voltage
125KV
Max. Dc output current
300mA
X-Ray tube
model
XD51-2040/125
focus
small focus 1X1mm big focus 2X2mm
Photographic bed(YZ-I)
Bed face(LXWXH)
2000X665X700mm
Moving range of bed face
640mm in length, 200mm in breadth
Ray-filter of bed
Travel in length >=500mm
Grid density N28
Convergence distance fo 100cm
Grid ratio R8
Pillar of X-ray source unit
Moving in length along photographic bed
1800mm or 1600mm
Distance around as moving up and down(along upright post)
650-1800mm
Rotary around the center of cross arm
± 180°
Rotary around the shaft line of X-ray annular tubes
-10 ° - 60° - +120°
Max size of cassette for photograph
356mmX432mm(14”X17”)